2007-2. How Little Can I See?

With trace analysis, a lower limit of detection is always welcomed. LCGC 25(2) 142-152 (2007) John W. Dolan Errata: There is an error in equation 1 of this month’s column. See LCGC 26(5) (2008) for complete explanation. In the November installment of “LC Troubleshooting” [LCGC 24(11) 1184-1190 (2006)], the y […]

Read More →



2002-9. Performance Qualification of LC Systems

Chromatographers can use a simple set of tests to check system performance. LCGC 20(9) 842-848 (2002) Gerald Hall and John W. Dolan Note that there are several errors in this document. The y-axis scale on Figure 1 should be microvolts, not millivolts. The calculations for Figure 1 are OK, but […]

Read More →

1998-10. Autosampler Precision

Several variables can affect autosampler injection precision and may prevent the instrument from passing performance tests LCGC 16(10) 910-914 (1998) John W. Dolan

Read More →